Arca
Skip to content
Tools
Tools

Tool, StripperW/BLD, 10S

Recall Z-2173-2015, part of FDA recall event 71534. See American Medical Systems Innovation Center - Silicon Valley's company page.

American Medical Systems Innovation Center - Silicon Valley started this Class II device recall on May 28, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Jan 19, 2016.

  • Classification: Class II
  • Started: May 28, 2015
  • Status: Terminated
  • Products in this event: 25

Why It Was Recalled

Class II

validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products

Record

Product
Tool, StripperW/BLD, 10S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective jacket off the end of an Endostat or SureFlex Fiber. The Fiber Stripper does not come in to direct contact with the patient. It is available in several sizes, corresponding to the Endostat or SureFlex Fiber core diameter size.
Quantity
16,710 total affected devices
Lots and codes
Model 80-10001-002; All product manufactured from April 8, 2012 to April 8, 2015
Distribution
Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.
Recalling firm
American Medical Systems Innovation Center - Silicon Valley, San Jose, CA, United States
How the firm notified
Letter
FDA classified
Jul 21, 2015

Other Products in This Recall

Recall event 71534; the first 24