Arca
Skip to content
Tools
Tools

Autopen Classic Automatic insulin delivery pen

Recall Z-2082-2015, part of FDA recall event 70783. See Owen Mumford USA's company page.

Owen Mumford USA, Inc. started this Class II device recall on Dec 11, 2014 (voluntary: firm initiated); FDA lists it as terminated, terminated Jul 11, 2017.

  • Classification: Class II
  • Started: Dec 11, 2014
  • Status: Terminated

Why It Was Recalled

Class II

The dials in the Autopens are not working - the dial does not stay dialed, but springs right back to 0. As a result, the patient could not dial up a dose of medicine to deliver via an injection.

Record

Product
Autopen Classic Automatic insulin delivery pen, 3.0ml x 2 unit 2-42 units, AN3800, Lot 7JH, NDC 08470-3800-01, The product is a non-sterile automatic subcutaneous injection device which delivers insulin.
Quantity
225 units
Lots and codes
Lot 7JH, UPC Code 384703800014
Distribution
Nationwide Distribution including FL, IL, NV, NJ, OH, and TN.
Recalling firm
Owen Mumford USA, Inc., Marietta, GA, United States
How the firm notified
Letter
FDA classified
Jul 15, 2015

Owen Mumford USA, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.