Unifine Pentips 5mm x 31G pen needles
Recall Z-0382-2018, part of FDA recall event 78483. See Owen Mumford USA's company page.
Owen Mumford USA, Inc. started this Class II device recall on Nov 6, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Feb 12, 2021.
- Classification: Class II
- Started: Nov 6, 2017
- Status: Terminated
- Products in this event: 2
Device may potentially be compromised by water in end user packaging. Exposure to liquid may compromise sterility. This may appear as discoloring or yellowing of the device's protective seal.
- Product
- Unifine Pentips 5mm x 31G pen needles, packaged in 30 ct pen needles. Product Usage: Sterile, non-toxic, non-pyrogenic, single use pen needles for the subcutaneous injection of drugs from pen injectors.
- Quantity
- 708 cartons (21,240 pen needles)
- Lots and codes
- UPC #384701150012: Batch No. AN 1150, Lot No.1705711.T098; Batch No. 1150REC, Lot No. 1705722.T099; Batch No. AN 1150HEB, Lot No. 1705722.T100.
- Distribution
- US Nationwide Distribution in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, MA, MI, MN, MO, MS, NC, NH NJ, NY, OH, OK, PA, TN, NX, UT, VA, VT, WA, WV
- Recalling firm
- Owen Mumford USA, Inc., Marietta, GA, United States
- How the firm notified
- Letter
- FDA classified
- Jan 17, 2018
- Unifine Pentips 5mm x 31G pen needles, packaged in 100 ct pen needles. Product Usage: Sterile, non-toxic,…Z-0381-2018 · Class II · Nov 6, 2017 · Terminated
Newest first, one product per recall event; search them all
- Unifine Pentips Plus, hypodermic needle - Product Usage: Pen needles with built in pen needle remover.Z-2078-2021 · Class II · May 26, 2021 · Ongoing
- Petfine Auto Injector, REF 3310VETZ-0264-2020 · Class II · Jul 17, 2019 · Ongoing
- Autopen Classic Automatic insulin delivery pen, 3.0ml x 2 unit 2-42 units, AN3800, Lot 7JH, NDC 08470-3800-01, The…Z-2082-2015 · Class II · Dec 11, 2014 · Terminated
- Autoject E1 Fixed needle device, Model Number AJ1310, UPC Code 384701310010 The Autoject E1 is an injector used by…Z-1366-2015 · Class II · Nov 24, 2014 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports