Alaris System PC Unit Model 8015 with software versions 9.17 and 9.19; Central…
Recall Z-2064-2016, part of FDA recall event 74142. See CareFusion 303's company page.
CareFusion 303, Inc. started this Class II device recall on May 12, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 24, 2018.
- Classification: Class II
- Started: May 12, 2016
- Status: Terminated
A patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110.
- Product
- Alaris System PC Unit Model 8015 with software versions 9.17 and 9.19; Central programming, monitoring and power supply component for the Alaris System.
- Quantity
- 69,693 units
- Lots and codes
- Serial no. 14566816 14566942 14566976 14567039 14567041 14567050 14567060 14567079 14567084 14567085 14567126 14567131 14567132 14567144 14412839 14509419 14509442 14509601 14510079 14510087 14517114 14517126 14517135 14517137 14517139 14517179 14517183 14517186 14517193 14517202 14517205 14517208 14517214 14517218 14517220 14517246 14517250 14517253 14517254 14517256 14517257 14517259 14517260 14517282 14517286 14517287 14517296 14517297 14517301 14517303 14517305 14517309 14517310 14517312 14517314 14517315 14517316 14517318 14517319 14517323 14517324 14517326 14517327 14517336 14517337…
- Distribution
- Distributed US (nationwide) and in Canada.
- Recalling firm
- CareFusion 303, Inc., San Diego, CA, United States
- How the firm notified
- Letter
- FDA classified
- Jun 28, 2016
Newest first, one product per recall event; search them all
- BD Pyxis MedStation ES Tower REF: 352 Medication cabinetZ-2302-2026 · Class II · Apr 30, 2026 · Ongoing
- BD Alaris" System with Guardrails" Suite MX with Point of Care UnitZ-1645-2026 · Class II · Feb 11, 2026 · Ongoing
- BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03,…Z-1028-2026 · Class II · Dec 3, 2025 · Ongoing
- BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES…Z-0962-2026 · Class II · Nov 20, 2025 · Ongoing
- BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a…Z-1217-2026 · Class II · Nov 6, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports