Arca
Skip to content
Tools
Tools

PROBEAT-V, software versions 4.8.1S

Recall Z-1892-2019, part of FDA recall event 83056. See Hitachi America, Ltd., Power Systems Division's company page.

Hitachi America, Ltd., Power Systems Division started this Class II device recall on Apr 15, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 24, 2020.

  • Classification: Class II
  • Started: Apr 15, 2019
  • Status: Terminated

Why It Was Recalled

Class II

There is a potential for a discrepant target position when using 3D3D matching mode in PIAS (Positioning Image Analysis System) software installed in the PROBEAT-V, proton therapy systems.

Record

Product
PROBEAT-V, software versions 4.8.1S, and 5.1M Radiation therapy system.
Quantity
3
Lots and codes
2015B01, A01, A02
Distribution
Distributed to accounts located in AZ, MN, and TN. Foreign distribution to Japan.
Recalling firm
Hitachi America, Ltd., Power Systems Division, Houston, TX, United States
How the firm notified
Letter
FDA classified
Jun 27, 2019

Hitachi America, Ltd., Power Systems Division's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc