PROBEAT WITH DISCRETE SPOT SCANNING SYSTEM Product Usage: Hitachi PROBEAT with DSSS is a…
Recall Z-1838-2015, part of FDA recall event 71056. See Hitachi America, Ltd., Power Systems Division's company page.
Hitachi America, Ltd., Power Systems Division started this Class II device recall on Apr 21, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 24, 2016.
- Classification: Class II
- Started: Apr 21, 2015
- Status: Terminated
At a PROBEAT III site installed in Japan, the operator of the system found that the radiation was delivered inconsistent with the treatment plan and stopped the irradiation. The irradiation control system may be potentially affected when rebooting a part of the control system.
- Product
- PROBEAT WITH DISCRETE SPOT SCANNING SYSTEM Product Usage: Hitachi PROBEAT with DSSS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
- Quantity
- 3
- Lots and codes
- No serial number or lot code. Each system is designed for a specific facility therefore no codes are needed for traceability
- Distribution
- Distribution in US (Texas) and Japan
- Recalling firm
- Hitachi America, Ltd., Power Systems Division, Houston, TX, United States
- How the firm notified
- Letter
- FDA classified
- Jul 17, 2015