Apo B Reagent, REF: OSR6143
Recall Z-1815-2025, part of FDA recall event 96746. See Beckman Coulter Ireland's company page.
Beckman Coulter Ireland, Inc. started this Class III device recall on Apr 23, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class III
- Started: Apr 23, 2025
- Status: Ongoing
The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.
- Product
- Apo B Reagent, REF: OSR6143
- Quantity
- 2747 units
- Lots and codes
- Lot # 2631/UDI: 15099590010409
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MD, MI, NJ, NY, OH, OK, OR, PA, TX, VA, WA, WV and the countries of Argentina, Australia, Belarus, Belgium, Bulgaria, Canada, China, Czechia, Egypt, Estonia, France, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Korea, Republic of, Malaysia, Netherlands, New Zealand, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam.
- Recalling firm
- Beckman Coulter Ireland, Inc., O'Callaghan'S Mills, N/A, Ireland
- How the firm notified
- Letter
- FDA classified
- May 23, 2025
- The LDL-Cholesterol assay is comprised of two distinct phases. In phase one a unique detergent solubilizes cholesterol…Z-1135-2026 · Class II · Dec 8, 2025 · Ongoing
- BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system reagent for the…Z-0997-2026 · Class II · Oct 14, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports