Arca
Skip to content
Tools
Tools

BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system…

Recall Z-0997-2026, part of FDA recall event 97842. See Beckman Coulter Ireland's company page.

Beckman Coulter Ireland, Inc. started this Class II device recall on Oct 14, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Oct 14, 2025
  • Status: Ongoing

Why It Was Recalled

Class II

Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.

Record

Product
BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system reagent for the quantitative determination of Bicarbonate in human serum and plasma on Beckman Coulter AU analyzers.
Quantity
36,981 reagents (U.S.-35,320, O.U.S.-1,661)
Lots and codes
Catalog Numbers / UDI-DI codes: OSR6137 - 15099590020187 Lot Numbers: 2681 2682 2683 2684 2685 2686 2687 2688 2689 2690 2693 2694 2695 2696 OSR6237 - 15099590010348 Lot Numbers: 2702 2703 2704 2705 2706 2707 2708 2709 2710 2712 2713 2714 2715 2716 OSR6637 - 15099590060206 Lot Numbers: 2726 2728 2729 2730 2731 2732 2733 2734 2736 2737 2738 2739
Distribution
Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guan, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Brazil, Canada, Ecuador, Ireland, Mexico, Netherlands, Panama, Taiwan, and Province of China.
Recalling firm
Beckman Coulter Ireland, Inc., O'Callaghan'S Mills, N/A, Ireland
How the firm notified
Letter
FDA classified
Dec 23, 2025

Beckman Coulter Ireland, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports