Arca
Skip to content
Tools
Tools

BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Matrix Drawers…

Recall Z-1785-2022, part of FDA recall event 90600. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class II device recall on May 24, 2022 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: May 24, 2022
  • Status: Ongoing

Why It Was Recalled

Class II

The automated dispensing cabinet software update included a .dll file that is not compatible with MedBank Cabinet stations with MedCart Matrix drawers. The cabinet system will stop functioning (Matrix drawers will stop opening in the MedBank Application) leading to delay in access to medications or supplies needed in residential or procedural care facilities.

Record

Product
BD Pyxis MedBank Cabinet Software v2.1.2.9 Legacy MedBank Cabinets with Matrix Drawers include: . CUBEX MedFlex operating with Windows 7 . MedBank 200 Main operating with Windows 7 . MedBank 500 Main operating with Windows 7. An automated dispensing cabinet intended to securely store and dispense medications to health care providers in a clinical setting.
Quantity
686 cabinets
Lots and codes
Software Version - v2.1.2.9 UDI Code: N/A Cabinet System Serial Numbers containing the affected software: 14473379 15268856 14728142 15080054 14891492 15216161 13819501 14443828 14443829 14443838 14461825 14461827 14461829 14461830 14461836 14461838 14461840 14461842 14461843 14461845 14461848 14473356 14473359 14473371 14473376 14473377 14491507 14491510 14491511 14491513 14491514 14491516 14491518 14491519 14491523 14491524 14491526 14491529 14491533 14491534 14491535 14491536 14491537 14509122 14509123 14509124 14509126 14509128 14509130 14509136 14509137 14509138 14509139 14509141…
Distribution
U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS,KY, LA, MA, MD, ME,. MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Sep 23, 2022

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc