Arca
Skip to content
Tools
Tools

Alaris Pump, Model 8100

Recall Z-1768-2019, part of FDA recall event 82143. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class I device recall on Apr 13, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 26, 2023.

  • Classification: Class I
  • Started: Apr 13, 2018
  • Status: Terminated

Why It Was Recalled

Class I

The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The separation of one or more bezel posts may result in free flow, over infusion, under infusion or interruption of infusion.

Record

Product
Alaris Pump, Model 8100, with bezel mechanical assemblies manufactured with FR-110 plastic resin. The following products are affected by this recall as they are pumps or pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic. Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N 10942012, P/N 49000007, and P/N 49000203), and Alaris Pump module Bezel Kit Assembly (P/N 10964559 and P/N 49000204). Product Usage: The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural. There are 6 bezel posts on the Alaris Pump bezel assembly.
Quantity
607618
Lots and codes
All pumps manufactured between April 2011 and June 2017, that have bezel mechanical assemblies manufactured with FR-110 plastic resin. And all pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic.
Distribution
Worldwide Distribution - U.S. Nationwide CA, TX, NC, AL, ID, TN, IN, MA, UT, HI, NH, PA, IL, OR, FL, MI, VA, AR, OH, GA, NY, MD, MN, NM, AZ, KY, WI, NE, MS, DC, MO, KS, SC, LA, MT, WA, NJ, CO, CT, OK, AK, AS, CU, DE, GU, IA, ME, MP, ND, NV, PR, SD, VT, WV, WY; OUS: AE, AU, BH, CA, CH, CN, GI, IL, IN, KW, MX, MY, NL, NZ, PH, PK, QA, SA, SG, TW, and ZA.
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Jul 16, 2019

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.