Catalys Precision Laser System-Catalys - U - US product; Catalys-1 - International…
Recall Z-1683-2015, part of FDA recall event 70672. See Optimedica's company page.
Optimedica Corporation started this Class II device recall on Feb 20, 2015 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 8, 2015.
- Classification: Class II
- Started: Feb 20, 2015
- Status: Terminated
Software anomaly on OptiMedica CATALYS System version 3.00.05 that may result in incorrect parameters provided for cataract incision surgeon templates and both eyes have the same parameters.
- Product
- Catalys Precision Laser System-Catalys - U - US product; Catalys-1 - International product; software version 3.00.05 The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.
- Quantity
- 171 distributed Worldwide
- Lots and codes
- Software version 3.00.05: Serial numbers - US: 44066413 44063113 44065413 44066213 44102714 44060012 44064813 44061012 44061812 44101614 44068314 44056612 44063713 44102614 44068914 44073514 44074814 44061612 44073714 44056312 44061412 44066613 44069414 44069814 44070914 44072714 44073014 44073114 44074514 44074614 44100714 44068014 44060612 44062813 44066914 44056012 44062213 44064513 44101014 44100814 44060312 44060412 44063513 44056712 44063313 44063913 44073314 44065613 44058312 44065013 44072114 44072414 44071914 44062613 44065213 44059912 44057312 44059712 44058212 44072814 44060712…
- Distribution
- Worldwide Distribution - US Nationwide, Germany, Switzerland, Australia, Spain, Japan, Great Britain, France, Netherlands, Austria, Singapore, South Korea, India, Brazil, Italy, Belgium, Egypt, Colombia, Hong Kong, and Canada.
- Recalling firm
- Optimedica Corporation, Sunnyvale, CA, United States
- How the firm notified
- Letter
- FDA classified
- May 28, 2015