Catalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser…

Recall Z-0649-2014, part of FDA recall event 67150. See Optimedica's company page.

Optimedica Corporation started this Class II device recall on Dec 20, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Jun 26, 2014.

  • Classification: Class II
  • Started: Dec 20, 2013
  • Status: Terminated

Why It Was Recalled

Class II

Potential for unexpected heating of the transformer mounting bolt located within the system chassis and the subsequent melting of a non-slip computer shelf mat, which may manifest an odor and visible smoke.

Record

Product
Catalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser System (International) - Catalys I 200-240V ~, 50/60 Hz, 15A, (2) 15A Resettable CB: Tested and complies with FCC Part 15 Class A. OptiMedica 1310 Moffett Park Drive Sunnyvale, CA 94089 The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.
Quantity
78 (39 in US, 39 internationally)
Lots and codes
All units shipped prior to Octotber 18, 2013. US Serial numbers: 44055311, 44055611, 44055811, 44056012, 44056312, 44056612, 44056712, 44056812, 44057312, 44057812, 44058012, 44058212, 44058312, 44059212, 44059312, 44059612, 44059712, 44059812, 44059912, 44060012, 44060112, 44060312, 44060412. International Serial numbers: 44000811, 44055211, 44055411, 44055711, 44055911, 44056112, 44056212, 44056412, 44056512, 44056912, 44057012, 44057112, 44057212, 44057412, 44057512, 44057612, 44057712, 44057912, 44058112, 44058512, 44058612, 44058912, 44059012. 44059112 44059412 44059512 44060212 44060512…
Distribution
Worldwide Distribution: US (nationwide) including states of: TX, NC, OH, IA, UT, KY, ME, CA, NY, MA. NJ, IL, PA,, TN, FL, MO, GA, HI, SC, NE, MI, and GA and countries of: Dominican Republic, Germany, Switzerland, Australia, Spain, Japan, United Kingdom, Canada, France, The Netherlands, Austria, and Singapore.
Recalling firm
Optimedica Corporation, Sunnyvale, CA, United States
How the firm notified
Letter
FDA classified
Jan 8, 2014

Optimedica Corporation's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.