Catalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser…
Recall Z-0649-2014, part of FDA recall event 67150. See Optimedica's company page.
Optimedica Corporation started this Class II device recall on Dec 20, 2013 (voluntary: firm initiated); FDA lists it as terminated, terminated Jun 26, 2014.
- Classification: Class II
- Started: Dec 20, 2013
- Status: Terminated
Potential for unexpected heating of the transformer mounting bolt located within the system chassis and the subsequent melting of a non-slip computer shelf mat, which may manifest an odor and visible smoke.
- Product
- Catalys Precision Laser System (United States) - Catalys-U: Catalys Precision Laser System (International) - Catalys I 200-240V ~, 50/60 Hz, 15A, (2) 15A Resettable CB: Tested and complies with FCC Part 15 Class A. OptiMedica 1310 Moffett Park Drive Sunnyvale, CA 94089 The OptiMedica Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens.
- Quantity
- 78 (39 in US, 39 internationally)
- Lots and codes
- All units shipped prior to Octotber 18, 2013. US Serial numbers: 44055311, 44055611, 44055811, 44056012, 44056312, 44056612, 44056712, 44056812, 44057312, 44057812, 44058012, 44058212, 44058312, 44059212, 44059312, 44059612, 44059712, 44059812, 44059912, 44060012, 44060112, 44060312, 44060412. International Serial numbers: 44000811, 44055211, 44055411, 44055711, 44055911, 44056112, 44056212, 44056412, 44056512, 44056912, 44057012, 44057112, 44057212, 44057412, 44057512, 44057612, 44057712, 44057912, 44058112, 44058512, 44058612, 44058912, 44059012. 44059112 44059412 44059512 44060212 44060512…
- Distribution
- Worldwide Distribution: US (nationwide) including states of: TX, NC, OH, IA, UT, KY, ME, CA, NY, MA. NJ, IL, PA,, TN, FL, MO, GA, HI, SC, NE, MI, and GA and countries of: Dominican Republic, Germany, Switzerland, Australia, Spain, Japan, United Kingdom, Canada, France, The Netherlands, Austria, and Singapore.
- Recalling firm
- Optimedica Corporation, Sunnyvale, CA, United States
- How the firm notified
- Letter
- FDA classified
- Jan 8, 2014
Newest first, one product per recall event; search them all
- Catalys Precision Laser System-Catalys - U - US product; Catalys-1 - International product; software version 3.00.05 …Z-1683-2015 · Class II · Feb 20, 2015 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports