Alaris PC unit, Model 8015 The Alaris PC unit is the central programming
Recall Z-1606-2016, part of FDA recall event 73631. See CareFusion 303's company page.
CareFusion 303, Inc. started this Class II device recall on Mar 24, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Jun 29, 2018.
- Classification: Class II
- Started: Mar 24, 2016
- Status: Terminated
The Alaris PC units model 8015 may display a system error code 133.6080 due to failure with the super capacitor (C245) at power up on the Alaris PC unit logic boards.
- Product
- Alaris PC unit, Model 8015 The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.
- Quantity
- 23,397 units
- Lots and codes
- Serial no. 13734283 13734308 13734309 13734310 13734312 13734313 13734337 13734339 13734345 13734351 13734352 13734353 13734354 13734355 13734361 13734364 13734365 13734366 13734367 13734368 13734370 13734374 13734376 13734377 13734379 13734382 13734383 13734408 13734413 13734414 13734415 13734418 13734424 13734427 13734430 13734431 13734432 13734435 13734437 13734440 13734441 13734452 13734453 13734454 13734455 13734456 13734457 13734458 13734459 13734460 13734461 13734462 13734463 13734464 13734465 13734466 13734467 13734468 13734469 13734470 13734474 13734476 13734479 13734486 13734487…
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Australia, Great Britain and Canada.
- Recalling firm
- CareFusion 303, Inc., San Diego, CA, United States
- How the firm notified
- Letter
- FDA classified
- May 4, 2016
Newest first, one product per recall event; search them all
- BD Pyxis MedStation ES Tower REF: 352 Medication cabinetZ-2302-2026 · Class II · Apr 30, 2026 · Ongoing
- BD Alaris" System with Guardrails" Suite MX with Point of Care UnitZ-1645-2026 · Class II · Feb 11, 2026 · Ongoing
- BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03,…Z-1028-2026 · Class II · Dec 3, 2025 · Ongoing
- BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES…Z-0962-2026 · Class II · Nov 20, 2025 · Ongoing
- BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a…Z-1217-2026 · Class II · Nov 6, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports