Arca
Skip to content
Tools
Tools

BD Pyxis MedStation ES (Med ES Main)

Recall Z-1579-2023, part of FDA recall event 92061. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class II device recall on Apr 5, 2023 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Apr 5, 2023
  • Status: Ongoing

Why It Was Recalled

Class II

Automated dispensing cabinet software is experiencing: 1) ES device download failure resulting in partial patient information loss that leads to 2) partial loss of patient transaction data, seen as a server upload processing error. Outdated/incorrect information display may contribute to the removal of medication to which the patient is allergic, incorrect dose amount, or discontinued medication.

Record

Product
BD Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES (Med ES Main Tower), REF: 352; BD Pyxis Anesthesia Station ES, REF: 327; BD Pyxis CII Safe ES Tower Main, REF: 1116-00; BD Pyxis CII Safe ES, Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01
Quantity
1,474
Lots and codes
REF/UDI-DI/Software: 323/10885403512667/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4; 352/10885403512674/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4; 327/10885403477836/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4; 1116-00/10885403512605/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4; 107-255-01/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4
Distribution
US Nationwide distribution in the states of KS, NY, NC, LA, MA, MI.
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
May 11, 2023

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports