Arca
Skip to content
Tools
Tools

Alaris System PC unit

Recall Z-1520-2017, part of FDA recall event 75353. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class II device recall on Nov 1, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 24, 2020.

  • Classification: Class II
  • Started: Nov 1, 2016
  • Status: Terminated

Why It Was Recalled

Class II

Reports where the Low Battery alarm and/or the Very Low Battery alarm are not being triggered before the battery is discharged and all infusion channels are stopped.

Record

Product
Alaris System PC unit, model no. 8000 and 8015. The central programming, monitoring and power supply component for the Alaris infusion pump System.
Quantity
613,800 total units (575,221 units in US)
Lots and codes
all units
Distribution
Worldwide distribution. US nationwide, Europe, Australia, New Zealand, South Africa, Greater Asia, Middle East, and Canada.
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Mar 21, 2017

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc