CareFusion Alaris Syringe Module
Recall Z-1361-2020, part of FDA recall event 84579. See CareFusion 303's company page.
CareFusion 303, Inc. started this Class I device recall on Feb 4, 2020 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class I
- Started: Feb 4, 2020
- Status: Ongoing
- Products in this event: 4
1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders
- Product
- CareFusion Alaris Syringe Module, Model 8110
- Quantity
- 114835
- Lots and codes
- Software versions 12.1.0, and 9.33 and prior
- Distribution
- Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN, AL, NH, OK, DC, NJ, MO, SD, UT, NM, MD, VT, MP, ND, AS; and countries of: Canada, Australia, New Zealand, the Middle East, South Africa, Taiwan, Southeast Asia and the European Union.
- Recalling firm
- CareFusion 303, Inc., San Diego, CA, United States
- How the firm notified
- Letter
- FDA classified
- Mar 5, 2020
- CareFusion Alaris System, Alaris PC unit models 8000 and 8015 (UDI: 10885403801518)Z-1359-2020 · Class I · Feb 4, 2020 · Ongoing
- CareFusion Alaris PCA Module, Model 8120Z-1362-2020 · Class I · Feb 4, 2020 · Ongoing
- CareFusion Alaris Pump Module, Model 8100Z-1360-2020 · Class I · Feb 4, 2020 · Ongoing
- BD Pyxis MedStation ES Tower REF: 352 Medication cabinetZ-2302-2026 · Class II · Apr 30, 2026 · Ongoing
- BD Alaris" System with Guardrails" Suite MX with Point of Care UnitZ-1645-2026 · Class II · Feb 11, 2026 · Ongoing
- BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03,…Z-1028-2026 · Class II · Dec 3, 2025 · Ongoing
- BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES…Z-0962-2026 · Class II · Nov 20, 2025 · Ongoing
- BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a…Z-1217-2026 · Class II · Nov 6, 2025 · Ongoing