Amicus Exchange Kit - Product Usage: intended for use in the collection of blood…
Recall Z-1344-2021, part of FDA recall event 87407. See Fenwal's company page.
Fenwal Inc started this Class II device recall on Feb 25, 2021 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 18, 2023.
- Classification: Class II
- Started: Feb 25, 2021
- Status: Terminated
- Products in this event: 3
Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.
- Product
- Amicus Exchange Kit - Product Usage: intended for use in the collection of blood components and mononuclear cells.
- Quantity
- 2466 units
- Lots and codes
- Product Code X6R2339; UDI: 04086000100151; Batch Numbers FA20F08147 (Exp. 06/07/2022), FA20H12145 (Exp. 08/11/2022), FA20J27157 (Exp. 10/26/2022)
- Distribution
- US Nationwide distribution in the states of GA, TX, TN, MD, CA, KS, MN, MA, NM, NC, LA, CO, KY, NV, PA, UT, MI, NJ, IL, MD, OH, NY, FL.
- Recalling firm
- Fenwal Inc, Lake Zurich, IL, United States
- How the firm notified
- Letter
- FDA classified
- Apr 2, 2021
- Amicus MNC Apheresis Kit - Double Needle - Product Usage: intended for use in the collection of blood components and…Z-1343-2021 · Class II · Feb 25, 2021 · Terminated
- Amicus Exchange Kit Therapeutics - Product Usage: intended for use in the collection of blood components and…Z-1345-2021 · Class II · Feb 25, 2021 · Terminated
Newest first, one product per recall event; search them all
- Amicus MNC Apheresis Kit Double Needle. Component of Amicus Separator multiprocedural apheresis platformZ-1254-2023 · Class II · Feb 14, 2023 · Ongoing
- VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC…Z-2064-2019 · Class I · Jun 24, 2019 · Terminated
- Fresenius Kabi CATSmart device (Continuous autotransfusion System). Labeled as the following kits: 1. AT-1 sets…Z-1209-2019 · Class II · Mar 5, 2019 · Terminated
- Vacuum Unit for CATSmart Vacuum Pump. Fresenius CATSmart (Continuous Autotransfusion System) is indicated for the…Z-1059-2019 · Class II · Feb 11, 2019 · Terminated
- AMICUS Exchange Kit; Product Usage: This kit is designed for use with the AMICUS separator for the Therapeutic…Z-1523-2013 · Class II · May 24, 2013 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports