Arca
Skip to content
Tools
Tools

Amicus Exchange Kit - Product Usage: intended for use in the collection of blood…

Recall Z-1344-2021, part of FDA recall event 87407. See Fenwal's company page.

Fenwal Inc started this Class II device recall on Feb 25, 2021 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 18, 2023.

  • Classification: Class II
  • Started: Feb 25, 2021
  • Status: Terminated
  • Products in this event: 3

Why It Was Recalled

Class II

Some lots of Amicus MNC Apheresis and Amicus Exchange kits have leaking centrifuge packs during a procedure on the Amicus Separator using therapeutic protocols only. The defect is a blood leak at the boot, elbow, or in the channel on the separation chamber of the centrifuge pack.

Record

Product
Amicus Exchange Kit - Product Usage: intended for use in the collection of blood components and mononuclear cells.
Quantity
2466 units
Lots and codes
Product Code X6R2339; UDI: 04086000100151; Batch Numbers FA20F08147 (Exp. 06/07/2022), FA20H12145 (Exp. 08/11/2022), FA20J27157 (Exp. 10/26/2022)
Distribution
US Nationwide distribution in the states of GA, TX, TN, MD, CA, KS, MN, MA, NM, NC, LA, CO, KY, NV, PA, UT, MI, NJ, IL, MD, OH, NY, FL.
Recalling firm
Fenwal Inc, Lake Zurich, IL, United States
How the firm notified
Letter
FDA classified
Apr 2, 2021

Other Products in This Recall

Recall event 87407

Fenwal Inc's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports