Fresenius Kabi CATSmart device (Continuous autotransfusion System). Labeled as the…
Recall Z-1209-2019, part of FDA recall event 82395. See Fenwal's company page.
Fenwal Inc started this Class II device recall on Mar 5, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Oct 5, 2020.
- Classification: Class II
- Started: Mar 5, 2019
- Status: Terminated
Based on internal investigations, Fresenius Kabi has identified a multifaceted set of events involving the CATSmart Systems, including the device, software, and disposable which may result in the centrifuge tubing line making contact with the washing chamber. The contact may result in excessive wear of the centrifuge tubing or in extreme cases a puncture of the tubing.
- Product
- Fresenius Kabi CATSmart device (Continuous autotransfusion System). Labeled as the following kits: 1. AT-1 sets and kits (9005104 AT-1 Autotransfusion set, 9108504 ATF 120 Fast Start Kit, and 9108494 ATF 40 Fast Start Kits); 2. AT-3 Autotransfusion sets and kits (Part 9005444 AT-3 set, 9104444 ATF-120 Fast Start Kits (AT-3), or 9108444 ATF-40 Fast Start Kits AT-3) Product Usage: The Fresenius Kabi CATSmart device (Continuous auto transfusion System) is indicated for the processing of autologous shed blood collected interaoperatively and postoperatively to obtain washed packed red blood cells for reinfusion. The CATSmart Auto transfusion System is designed to reduce the need for allogeneic blood transfusion by re-infusing a patient s own lost blood during surgery. The set is used as a disposable component of the autotransfusion system during surgery and/or treatment of trauma to remove waste and improve the safety of autologous re-infused blood (auto transfusion). This is a single-use device.
- Quantity
- 342 devices
- Lots and codes
- Product Code: 9005901; Lot Numbers beginning with GM or H. Serial Numbers: 6CTA0209, 6CTA0210, 6CTA0211, 6CTA0212, 6CTA0213, 6CTA0214, 6CTA0215, 6CTA0216, 6CTA0217, 6CTA0218, 6CTA0219, 6CTA0221, 6CTA0222, 6CTA0223, 6CTA0224, 6CTA0225, 6CTA0226, 6CTA0227, 6CTA0228, 6CTA0260, 6CTA0261, 6CTA0262, 6CTA0263, 6CTA0264, 6CTA0265, 6CTA0266, 6CTA0267, 6CTA0268, 6CTA0269, 6CTA0300, 6CTA0301, 6CTA0302, 6CTA0303, 6CTA0304, 6CTA0305, 6CTA0306, 6CTA0307, 6CTA0308, 6CTA0309, 7CTA0328, 7CTA0329, 7CTA0330, 7CTA0331, 7CTA0332, 7CTA0334, 7CTA0335, 7CTA0336, 7CTA0337, 7CTA0358, 7CTA0359, 7CTA0360, 7CTA0361,…
- Distribution
- Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, FL, GA, HI, ID, IL, IN, LA, MA, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, and PR, and country of Canada.
- Recalling firm
- Fenwal Inc, Lake Zurich, IL, United States
- How the firm notified
- FDA classified
- Apr 23, 2019
- Amicus MNC Apheresis Kit Double Needle. Component of Amicus Separator multiprocedural apheresis platformZ-1254-2023 · Class II · Feb 14, 2023 · Ongoing
- Amicus Exchange Kit Therapeutics - Product Usage: intended for use in the collection of blood components and…Z-1345-2021 · Class II · Feb 25, 2021 · Terminated
- VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC…Z-2064-2019 · Class I · Jun 24, 2019 · Terminated
- Vacuum Unit for CATSmart Vacuum Pump. Fresenius CATSmart (Continuous Autotransfusion System) is indicated for the…Z-1059-2019 · Class II · Feb 11, 2019 · Terminated
- AMICUS Exchange Kit; Product Usage: This kit is designed for use with the AMICUS separator for the Therapeutic…Z-1523-2013 · Class II · May 24, 2013 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports