Arca
Skip to content
Tools
Tools

Abutment Screw Retrieval Instrument 3.0/NP - Product Usage: The instruments for abutment…

Recall Z-1322-2020, part of FDA recall event 84450. See Nobel Biocare USA's company page.

Nobel Biocare Usa Llc started this Class II device recall on Mar 14, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Feb 15, 2024.

  • Classification: Class II
  • Started: Mar 14, 2019
  • Status: Terminated
  • Products in this event: 24

Why It Was Recalled

Class II

Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.

Record

Product
Abutment Screw Retrieval Instrument 3.0/NP - Product Usage: The instruments for abutment screw retrieval can be used if an abutment screw is broken and a part of the screw is left inside the implant screw channel. If the implant thread needs to be cleaned before placing a new screw, a screw tap repair can be used. The instruments are intended for both simple and advanced treatments.
Quantity
233 units
Lots and codes
(01)07332747084085(10)73585 (01)07332747084085(10)73430 Catalog # 37503
Distribution
Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.
Recalling firm
Nobel Biocare Usa Llc, Yorba Linda, CA, United States
How the firm notified
E-Mail
FDA classified
Feb 19, 2020

Other Products in This Recall

Recall event 84450

Nobel Biocare Usa Llc's Other Recalls

Newest first, one product per recall event; search them all