NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R
Recall Z-0658-2022, part of FDA recall event 89465. See Nobel Biocare USA's company page.
Nobel Biocare Usa Llc started this Class II device recall on Dec 17, 2021 (voluntary: firm initiated); FDA lists it as terminated, terminated Oct 25, 2024.
- Classification: Class II
- Started: Dec 17, 2021
- Status: Terminated
Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.
- Product
- NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R
- Quantity
- 702 pieces
- Lots and codes
- Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497
- Distribution
- Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.
- Recalling firm
- Nobel Biocare Usa Llc, Yorba Linda, CA, United States
- How the firm notified
- Letter
- FDA classified
- Feb 18, 2022
Newest first, one product per recall event; search them all
- NobelParallel Conical Connection NP, 3.75x10mm (Endosseous Dental Implant); REF 37965; UDI …Z-1851-2020 · Class II · Aug 28, 2019 · Terminated
- Snappy Abutment 5.5 CC RP 3 mm; pre-manufactured prosthetic components directly connected to the endosseous implant,…Z-1855-2020 · Class II · Aug 9, 2019 · Terminated
- NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619Z-2207-2020 · Class II · Jul 19, 2019 · Terminated
- Healing Abutment Conical Connection RP 0 3.6 x 5mm, REF: 36644, UDI: 07332747034462 - Product Usage: Dental implant…Z-2250-2020 · Class II · May 28, 2019 · Terminated
- Drill Tapered NP 3.5x11.5 mm Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills)…Z-1307-2020 · Class II · Mar 14, 2019 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports