Arca
Skip to content
Tools
Tools

NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R

Recall Z-0658-2022, part of FDA recall event 89465. See Nobel Biocare USA's company page.

Nobel Biocare Usa Llc started this Class II device recall on Dec 17, 2021 (voluntary: firm initiated); FDA lists it as terminated, terminated Oct 25, 2024.

  • Classification: Class II
  • Started: Dec 17, 2021
  • Status: Terminated

Why It Was Recalled

Class II

Due to 16 complaints related to drivers not engaging in implants. According to the firm's investigations, the incorrect offset gauge was potentially used during production of the implants. As a result, this could potentially lead to aborted or a delay in surgical procedure.

Record

Product
NobelReplace Conical Connection RP 4.3x11.5 mm REF 36707 STERILE R
Quantity
702 pieces
Lots and codes
Catalog Number: 36707 UDI Code: (01)07332747160529(10)13116495(11)210720(17)260618; (01)07332747160529(10)13116496(11)210720(17)260618; (01)07332747160529(10)13116497(11)210720(17)260618; Lot Numbers: 13116495, 13116496, 13116497
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, GA, HI, ID, IL, IN, KS, KY, MA, ME, MI, MN, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, VT, WA and WI. The countries of Canada, Mexico and Turkey.
Recalling firm
Nobel Biocare Usa Llc, Yorba Linda, CA, United States
How the firm notified
Letter
FDA classified
Feb 18, 2022

Nobel Biocare Usa Llc's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.