BD Alaris Infusion Pump Module Model 8100 Pump Module Keypad; and Pump Module Door…

Recall Z-1309-2021, part of FDA recall event 87387. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class I device recall on Mar 3, 2021 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class I
  • Started: Mar 3, 2021
  • Status: Ongoing

Why It Was Recalled

Class I

Pump Module keypad lifting, and Fluid ingress could result in 1) Unresponsive keys: module continues infusion, PC unit will not alarm and must be used for programming changes, may necessitate a different Pump Module. May result in infusion start delay/inability to titrate medication. 2) Stuck keys: PC unit alarms, module exhibits Channel Error, may result in infusion interruption/start delay

Record

Product
BD Alaris Infusion Pump Module Model 8100 Pump Module Keypad; and Pump Module Door Assembly Replacement Kit with Pump Module Keypad, Parts: 49000239; 49000346; 49000438; 49000439.
Quantity
160300
Lots and codes
Pump Module with Keypad: manufactured January 23, 2019 to December 5, 2019. Pump Module Door Assembly Kit with Keypad: Dated from January 15, 2019 to November 14, 2019.
Distribution
Worldwide distribution. US nationwide and countries of: AE, IL, IN, SA, BE, QA, SG, TW, KW, ZA, PH, AU, HU, NZ
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Apr 1, 2021