BD Pyxis: MedFlex, REF: 1119-00; MedBank Tower REF: 1145-00; MedBank MedPass
Recall Z-1206-2025, part of FDA recall event 96150. See CareFusion 303's company page.
CareFusion 303, Inc. started this Class II device recall on Jan 10, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jan 10, 2025
- Status: Ongoing
Automated dispensing device labeling is being updated to strengthen labeling around manual access to medications to address the risk of potential delays in retrieving contents/ accessing medications that may result from power loss or potential automated dispensing device malfunctions.
- Product
- BD Pyxis: MedFlex, REF: 1119-00; MedBank Tower REF: 1145-00; MedBank MedPass, REF: 139088-01 and associated BD Pyxis user/software/safety guides and quick reference guides.
- Quantity
- 1521
- Lots and codes
- User Guides: i. BD Pyxis MedBank Cabinet User Guide v2.4 ii. BD Pyxis MedBank MedPass User Guide v1.0 iii. BD Pyxis MedBank CR Station User Guide v3.10 iv. BD Pyxis MedBank myQLink Cloud Software User Guide v1.5 v. BD Pyxis Supply Cabinet Safety Guide (for SupplyBank models) Quick Reference Guides: i. BD Pyxis MedBank CR Station QRG (part number 155018-03) ii. BD Pyxis MedBank Cabinet QRG (PN 155023-03) iii. BD Pyxis MedBank RxNow CR Station QRG (PN 155026-03) iv. BD Pyxis MedBank RxNow Solution QRG 2.0 (PN 155027-02) v. BD Pyxis MedBank MedPass QRG (PN 155022-03) vi. BD Pyxis MedBank QFill…
- Distribution
- US Nationwide distribution including in the states of AZ, FL, NJ, KY, GA, CA, IL, MD, WI, OH, NC, MS, NY, MO, PA, OR, IN, TX, UT, OK, MI, WA, AL, NV, ME, WV, NE, DC, MA, MT, KS, CO, TN, IA, MN, AR, ID, SC, SD, CT, HI, VA, ND, AK, NM, VT, LA, WY.
- Recalling firm
- CareFusion 303, Inc., San Diego, CA, United States
- How the firm notified
- Letter
- FDA classified
- Feb 21, 2025
Newest first, one product per recall event; search them all
- BD Pyxis MedStation ES Tower REF: 352 Medication cabinetZ-2302-2026 · Class II · Apr 30, 2026 · Ongoing
- BD Alaris" System with Guardrails" Suite MX with Point of Care UnitZ-1645-2026 · Class II · Feb 11, 2026 · Ongoing
- BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03,…Z-1028-2026 · Class II · Dec 3, 2025 · Ongoing
- BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES…Z-0962-2026 · Class II · Nov 20, 2025 · Ongoing
- BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a…Z-1217-2026 · Class II · Nov 6, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports