SwishPlus Implant 5.7mmD x 8mmL
Recall Z-1126-2020, part of FDA recall event 84706. See Implant Direct Sybron Manufacturing's company page.
Implant Direct Sybron Manufacturing LLC started this Class II device recall on Jan 9, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 27, 2021.
- Classification: Class II
- Started: Jan 9, 2019
- Status: Terminated
Due to a complaint received and internal investigation it was determined that implant vials were distributed with incorrect color coded vials and caps. The part/lot numbers were correct, the customer received the product they ordered, however the color on the vial label and cap labels were incorrect.Specifically, the color coded vial and cap does not match the implant size described on the vial. As a result of the incorrect color on the vial and cap, the surgeon may not be certain if the implant received was the correct size. This could result in a delay in surgery, risk of infection if the surgical site has to be closed up and a second round of anesthesia.
- Product
- SwishPlus Implant 5.7mmD x 8mmL, SBM 6.5mmD Platform, Part Number 925708
- Quantity
- 11 units
- Lots and codes
- Lot # 109038 No affected devices distributed within the US.
- Distribution
- US: None OUS: Iceland, Sweden, Spain,
- Recalling firm
- Implant Direct Sybron Manufacturing LLC, Westlake Village, CA, United States
- How the firm notified
- Letter
- FDA classified
- Feb 10, 2020
- The Drill Stop Kit is the version of the Mini Sterilizable Tray that is used to store, sterilize, verify Drill Stop…Z-2172-2025 · Class II · May 1, 2025 · Ongoing
- LEGACY SMARTBASE, NE 3.0D 1MML, REF: 8730-81NE; LEGACY SMARTBASE, NE 3.0D 2MML, REF 8730-82NE; LEGACY SMARTBASE…Z-2037-2023 · Class II · May 2, 2023 · Ongoing
- ImplantDirect, REF 834711, Legacy 2 Implant 4.7mmD x 11.5mmL, 4.5mmD Platform,Z-0197-2022 · Class II · Sep 23, 2021 · Ongoing
- Implant Direct, CE 0086, REF: 855710, Legacy 3 Impant, Surface: SBM, Qty:1, Size:5.7mmD x 10mmL, 5.7mmD Platform,…Z-1157-2020 · Class II · Oct 11, 2019 · Terminated
- Legacy 3 Implant, Size: 4.2mmD x 13mmL, 3.5mmD Platform, REF: 864213Z-1041-2020 · Class II · Oct 7, 2019 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports