Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP…
Recall Z-1037-2026, part of FDA recall event 98091. See Integra LifeSciences's company page.
Integra LifeSciences Corp. (NeuroSciences) started this Class II device recall on Dec 12, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Dec 12, 2025
- Status: Ongoing
- Products in this event: 2
Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.
- Product
- Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the CereLink ICP Sensor (ICP Sensor) and a 14-gauge Tuohy needle with stylet. The ICP Sensor is a nylon tube with a microminiature strain gauge pressure transducer (sensing element) mounted at one end and an electrical connector at the other end. It is designed for use with a Codman intracranial pressure monitoring device. The ICP Sensor Basic Kit is designed for use with the Codman Cranial Hand Drill. The drill facilitates access to the intraparenchymal area. The drill is also available as a component of the Codman Cranial Access Kit. Use of the ICP Sensor Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.
- Quantity
- 5,116 units
- Lots and codes
- Catalog Number: 826850. UDI-DI: 10381780520672. Lot Numbers: 7261958, 7266691, 7266692, 7288060, 7293837, 7293836, 7306040, 7306041, 7306042, 7306043, 7306044, 7306045, 7306046, 7306047, 7306050, 7306051, 7306052, 7307670, 7307671, 7307672, 7325482, 7307675, 7325481, 7325483, 7325484, 7307673, 7307674, 7325485, 7325486, 7336611, 7336612, 7336614, 7336623, 7336613, 7336615, 7336616, 7336618, 7336619, 7336620, 7336621, 7336622, 7306048, 7306049, 7336626, 7336627, 7336629, 7347166, 7336628, 7346295, 7346297, 7346299, 7346301, 7365434, 7365435, 7392019, 7396607, 7396608, 7396609, 7396610,…
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
- Recalling firm
- Integra LifeSciences Corp. (NeuroSciences), Princeton, NJ, United States
- How the firm notified
- Letter
- FDA classified
- Jan 8, 2026
- Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when…Z-1036-2026 · Class II · Dec 12, 2025 · Ongoing
Newest first, one product per recall event; search them all
- OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.Z-2714-2026 · Class II · Jun 5, 2026 · Ongoing
- Codman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD…Z-2173-2026 · Class II · Apr 10, 2026 · Ongoing
- Cytal Wound Matrix 2-Layer 10x15 cm. Product ID: WSM1015.Z-1818-2026 · Class II · Mar 19, 2026 · Ongoing
- CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.Z-1828-2026 · Class II · Mar 6, 2026 · Ongoing
- MEDIHONEY CALCIUM ALGINATE WITH ACTIVE LEPTOSPERMUM HONEY. Model Numbers: (1) 31012, (2) 31022, (3) 31045. …Z-1364-2026 · Class II · Jan 16, 2026 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports