OmniTract Surgical Retractor System
Recall Z-2714-2026, part of FDA recall event 99193. See Integra LifeSciences's company page.
Integra LifeSciences Corp. (NeuroSciences) started this Class II device recall on Jun 5, 2026 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Jun 5, 2026
- Status: Ongoing
The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
- Product
- OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.
- Quantity
- 633 units
- Lots and codes
- Product Code 3032, UDI 10381780268468; Product Code 3055, UDI 10381780268505; Product Code 3069, UDI 10381780268512; Product Code 3076, UDI 10381780268529; Product Code 3086, UDI 10381780268536; Product Code 3092, UDI 10381780268550; Product Code 3093, UDI 10381780268567; Product Code 3096, UDI 10381780268581; Product Code 3098, UDI 10381780268598; Product Code 3099, UDI 10381780268604; Product Code 3133, UDI 10381780268680; Product Code 3137, UDI 10381780268697; Product Code 3142, UDI 10381780268710; Product Code 3146, UDI 10381780268734; Product Code 3148, UDI 10381780268758; Product Code…
- Distribution
- Worldwide distribution - US Nationwide including Puerto Rico, and the countries of Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brunei Darussalam, Bulgaria, Canada, Chile, China, Czechia, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Lithuania, Malaysia, Malta, Mauritius, Mexico, Nepal, Netherlands, New Zealand, Norway, Pakistan, State of Palestine, Philippines, Poland, Portugal, Qatar, Republic of North Macedonia, Reunion, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan (Province of China), Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, and Viet Nam.
- Recalling firm
- Integra LifeSciences Corp. (NeuroSciences), Princeton, NJ, United States
- How the firm notified
- Letter
- FDA classified
- Jul 9, 2026
Newest first, one product per recall event; search them all
- Codman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD…Z-2173-2026 · Class II · Apr 10, 2026 · Ongoing
- Cytal Wound Matrix 2-Layer 10x15 cm. Product ID: WSM1015.Z-1818-2026 · Class II · Mar 19, 2026 · Ongoing
- CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.Z-1828-2026 · Class II · Mar 6, 2026 · Ongoing
- MEDIHONEY CALCIUM ALGINATE WITH ACTIVE LEPTOSPERMUM HONEY. Model Numbers: (1) 31012, (2) 31022, (3) 31045. …Z-1364-2026 · Class II · Jan 16, 2026 · Ongoing
- Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when…Z-1036-2026 · Class II · Dec 12, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports