BD Pyxis MedBank System - Product Label/labeling pending
Recall Z-0928-2023, part of FDA recall event 91099. See CareFusion 303's company page.
CareFusion 303, Inc. started this Class II device recall on Dec 12, 2022 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Dec 12, 2022
- Status: Ongoing
Due to customer complaints related to the MedBank software which may indicate that a different medication is loaded in a cubie (drawer) pocket than what is physically loaded and listed on the pocket label. As a result of this issue, a patient could receive the wrong medication or wrong dose of medication that could lead to patient harm.
- Product
- BD Pyxis MedBank System - Product Label/labeling pending
- Quantity
- 6,796 systems
- Lots and codes
- BD Pyxis" MedBank Tower Main UDI Code: 10885403512568 Catalog Numbers: 169-23 to 169-81, 169-100, 169-132, 169-138, 169-139 BD Pyxis" MedBank Mini Main UDI Code: 10885403512544 Catalog Numbers: 169-101 - 169-113, 169-137 BD Pyxis" MEDFLEX BD Pyxis" MEDFLEX 2.0 UDI Codes:10885403512704 Catalog Numbers: 1119-00 1139-00 MedBank 200/500 UDI Code: N/A Catalog Numbers: 1137-00 BD Pyxis MEDPASS v1.0 or prior UDI Code: 10885403513817 Catalog Number: Not Provided BD Pyxis" MedBank Cabinet Software v2.2 or prior UDI Codes: 10885403513794 Catalog Number: N/A Affected Serial Numbers: 12940473 12971999…
- Distribution
- U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV.
- Recalling firm
- CareFusion 303, Inc., San Diego, CA, United States
- How the firm notified
- Letter
- FDA classified
- Jan 10, 2023
Newest first, one product per recall event; search them all
- BD Pyxis MedStation ES Tower REF: 352 Medication cabinetZ-2302-2026 · Class II · Apr 30, 2026 · Ongoing
- BD Alaris" System with Guardrails" Suite MX with Point of Care UnitZ-1645-2026 · Class II · Feb 11, 2026 · Ongoing
- BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03,…Z-1028-2026 · Class II · Dec 3, 2025 · Ongoing
- BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES…Z-0962-2026 · Class II · Nov 20, 2025 · Ongoing
- BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a…Z-1217-2026 · Class II · Nov 6, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports