Arca
Skip to content
Tools
Tools

BD Pyxis MedBank System - Product Label/labeling pending

Recall Z-0928-2023, part of FDA recall event 91099. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class II device recall on Dec 12, 2022 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Dec 12, 2022
  • Status: Ongoing

Why It Was Recalled

Class II

Due to customer complaints related to the MedBank software which may indicate that a different medication is loaded in a cubie (drawer) pocket than what is physically loaded and listed on the pocket label. As a result of this issue, a patient could receive the wrong medication or wrong dose of medication that could lead to patient harm.

Record

Product
BD Pyxis MedBank System - Product Label/labeling pending
Quantity
6,796 systems
Lots and codes
BD Pyxis" MedBank Tower Main UDI Code: 10885403512568 Catalog Numbers: 169-23 to 169-81, 169-100, 169-132, 169-138, 169-139 BD Pyxis" MedBank Mini Main UDI Code: 10885403512544 Catalog Numbers: 169-101 - 169-113, 169-137 BD Pyxis" MEDFLEX BD Pyxis" MEDFLEX 2.0 UDI Codes:10885403512704 Catalog Numbers: 1119-00 1139-00 MedBank 200/500 UDI Code: N/A Catalog Numbers: 1137-00 BD Pyxis MEDPASS v1.0 or prior UDI Code: 10885403513817 Catalog Number: Not Provided BD Pyxis" MedBank Cabinet Software v2.2 or prior UDI Codes: 10885403513794 Catalog Number: N/A Affected Serial Numbers: 12940473 12971999…
Distribution
U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, WA, WI, and WV.
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Jan 10, 2023

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all