Arca
Skip to content
Tools
Tools

BD Pyxis CII Safe ES

Recall Z-0672-2025, part of FDA recall event 95643. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class II device recall on Nov 14, 2024 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Nov 14, 2024
  • Status: Ongoing
  • Products in this event: 3

Why It Was Recalled

Class II

Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.

Record

Product
BD Pyxis CII Safe ES, REF: 1116-00
Quantity
218
Lots and codes
All software versions. REF/UDI-DI: 1116-00/10885403512605/ 15938609 15938608 15960426 15977244 15977243 15997587 15997584 15997586 16001276 16004167 16004170 16004168 16006959 16016028 16018286 16018285 16027068 16034363 16038606 16038608 16042040 16043100 16051041 16052300 16056892 16081629 16089006 16107354 16121844 16124176 16124177 16124175 16124174 16124178 16142019 16142717 16147368 16150604 16150317 16154380 16154381 16179910 16185906 16186402 16186403 16188916 16188917 16189876 16191858 16191855 16192478 16192476 16198735 16205805 16205708 16210899 16211121 16211303 16211474 16216310…
Distribution
Worldwide - US Nationwide distribution including in the states of OK, LA, FL, AR, VA, KY, PA, AL, NC, CA, MN, AK, TX, MI, ME, OH, OR, NY, IL, RI, GA, WI, NH, KS, CO, WA, MO, MD, AZ, SC, NV, WV, NM, IN, NJ, TN, DE, MT, MA, VT, DC, NE, CT, ID, WY, MS, UT, HI, IA, SD, ND, GU, MP, PR and the countries of CA, BR, JP, ES, NZ, AU, FR, AR, MX, IT, CL, TR, SA, MY, CH, IE, TH, SG, KW, BE, DE, PT, QA, MA, FI, AE, EG, BS, GB, BM, GR, BH, KR, MC, PL, LB, HK, IN, IL, MO, IQ, TN, NO, PH, CO, TW, CN, JO, OM.
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Dec 11, 2024

Other Products in This Recall

Recall event 95643

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.