C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Recall Z-0640-2020, part of FDA recall event 84216. See Integra LifeSciences's company page.

Integra LifeSciences Corp. started this Class II device recall on Oct 21, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated May 7, 2020.

  • Classification: Class II
  • Started: Oct 21, 2019
  • Status: Terminated
  • Products in this event: 25

Why It Was Recalled

Class II

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Record

Product
C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System
Quantity
N/A
Lots and codes
Danish OPS Manual P/N 60905209
Distribution
US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK
Recalling firm
Integra LifeSciences Corp., Plainsboro, NJ, United States
How the firm notified
Letter
FDA classified
Dec 8, 2019

Other Products in This Recall

Recall event 84216; the first 24

Integra LifeSciences Corp.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.