C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Recall Z-0613-2020, part of FDA recall event 84216. See Integra LifeSciences's company page.

Integra LifeSciences Corp. started this Class II device recall on Oct 21, 2019 (voluntary: firm initiated); FDA lists it as terminated, terminated May 7, 2020.

  • Classification: Class II
  • Started: Oct 21, 2019
  • Status: Terminated
  • Products in this event: 25

Why It Was Recalled

Class II

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Record

Product
C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System
Quantity
820
Lots and codes
US Ops Manual P/N 60905769
Distribution
US: AL AR AZ CA CO CT FL GA IA ID IL IN KS KY LA MA MD MI MN MO MS NC NE NJ NY OH OK OR PA SC TN UT VA WA WI WV International: SG MX CA JP AU CO TH HK
Recalling firm
Integra LifeSciences Corp., Plainsboro, NJ, United States
How the firm notified
Letter
FDA classified
Dec 8, 2019

Other Products in This Recall

Recall event 84216; the first 24

Integra LifeSciences Corp.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.