TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number…
Recall Z-0420-2026, part of FDA recall event 97677. See Integra LifeSciences's company page.
Integra LifeSciences Corp. (NeuroSciences) started this Class II device recall on Oct 3, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Oct 3, 2025
- Status: Ongoing
It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.
- Product
- TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01;
- Quantity
- 1198 units
- Lots and codes
- Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev 00 or Rev 01; UDI-DI: 10846835021103;
- Distribution
- US Nationwide distribution.
- Recalling firm
- Integra LifeSciences Corp. (NeuroSciences), Princeton, NJ, United States
- How the firm notified
- Letter
- FDA classified
- Nov 4, 2025
Newest first, one product per recall event; search them all
- OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.Z-2714-2026 · Class II · Jun 5, 2026 · Ongoing
- Codman CerebroFlo EVD Catheter with Endexo Technology. Contains 5 sets. Product ID: 37.550.501. The CerebroFlo EVD…Z-2173-2026 · Class II · Apr 10, 2026 · Ongoing
- Cytal Wound Matrix 2-Layer 10x15 cm. Product ID: WSM1015.Z-1818-2026 · Class II · Mar 19, 2026 · Ongoing
- CUSA Clarity C7000 Console, Software Version: 2.1.1.909. Ultrasonic surgical aspirator system.Z-1828-2026 · Class II · Mar 6, 2026 · Ongoing
- MEDIHONEY CALCIUM ALGINATE WITH ACTIVE LEPTOSPERMUM HONEY. Model Numbers: (1) 31012, (2) 31022, (3) 31045. …Z-1364-2026 · Class II · Jan 16, 2026 · Ongoing
- Codman Microsensor Basic Kit. Catalog Number: 626631US. Use of the CODMAN MICROSENSOR Basic Kit is indicated when…Z-1036-2026 · Class II · Dec 12, 2025 · Ongoing
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports