Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software…
Recall Z-0416-2026, part of FDA recall event 97686. See Spacelabs Healthcare's company page.
Spacelabs Healthcare, Ltd. started this Class II device recall on Sep 16, 2025 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Sep 16, 2025
- Status: Ongoing
Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
- Product
- Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.
- Quantity
- 142
- Lots and codes
- Catalog Number: 92810 UDI-DI code; (01)10841522125751 Software version: 11.6.0
- Distribution
- Worldwide - US Nationwide distribution in the state of MN and the countries of CA, DE, FR, GB, IT, NC.
- Recalling firm
- Spacelabs Healthcare, Ltd., Hertford, N/A, United Kingdom
- How the firm notified
- Letter
- FDA classified
- Oct 31, 2025
Newest first, one product per recall event; search them all
- Eclipse Mini REF 98900 The Eclipse MINI Model 98900 is a portable non-invasive continuous ambulatory ECG patch…Z-0685-2025 · Class II · Nov 25, 2024 · Ongoing
- Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming technology and is…Z-0563-2021 · Class II · Nov 6, 2020 · Terminated
- Arkon Anesthesia Delivery System with Arkon Software Version 2.70, Display Unit Assembly 650-1769-00 installed, and …Z-2589-2018 · Class I · Jul 11, 2018 · Terminated
- Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The…Z-0072-2018 · Class I · Oct 11, 2017 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports