Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software…

Recall Z-0416-2026, part of FDA recall event 97686. See Spacelabs Healthcare's company page.

Spacelabs Healthcare, Ltd. started this Class II device recall on Sep 16, 2025 (voluntary: firm initiated); FDA lists it as ongoing.

  • Classification: Class II
  • Started: Sep 16, 2025
  • Status: Ongoing

Why It Was Recalled

Class II

Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record

Record

Product
Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.
Quantity
142
Lots and codes
Catalog Number: 92810 UDI-DI code; (01)10841522125751 Software version: 11.6.0
Distribution
Worldwide - US Nationwide distribution in the state of MN and the countries of CA, DE, FR, GB, IT, NC.
Recalling firm
Spacelabs Healthcare, Ltd., Hertford, N/A, United Kingdom
How the firm notified
Letter
FDA classified
Oct 31, 2025