Arkon Anesthesia Delivery System
Recall Z-0072-2018, part of FDA recall event 78330. See Spacelabs Healthcare's company page.
Spacelabs Healthcare, Ltd. started this Class I device recall on Oct 11, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated Mar 12, 2018.
- Classification: Class I
- Started: Oct 11, 2017
- Status: Terminated
Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
- Product
- Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician
- Quantity
- 110 units
- Lots and codes
- UDI: 10841522100017
- Distribution
- Nationwide Distribution to the states of NM, WY, FL, NC, MS, CO, CT, ME, and AL.
- Recalling firm
- Spacelabs Healthcare, Ltd., Hertford, N/A, United Kingdom
- How the firm notified
- Letter
- FDA classified
- Nov 21, 2017
Newest first, one product per recall event; search them all
- Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product…Z-0416-2026 · Class II · Sep 16, 2025 · Ongoing
- Eclipse Mini REF 98900 The Eclipse MINI Model 98900 is a portable non-invasive continuous ambulatory ECG patch…Z-0685-2025 · Class II · Nov 25, 2024 · Ongoing
- Sentinel V10.x & V11.x, Models: 98200 & 98201 - Product Usage: designed using the latest programming technology and is…Z-0563-2021 · Class II · Nov 6, 2020 · Terminated
- Arkon Anesthesia Delivery System with Arkon Software Version 2.70, Display Unit Assembly 650-1769-00 installed, and …Z-2589-2018 · Class I · Jul 11, 2018 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports