MEVION S250 Product Usage: Proton Radiation Therapy System

Recall Z-0411-2017, part of FDA recall event 75494. See Mevion Medical Systems's company page.

Mevion Medical Systems, Inc. started this Class II device recall on Oct 20, 2016 (voluntary: firm initiated); FDA lists it as terminated, terminated May 15, 2017.

  • Classification: Class II
  • Started: Oct 20, 2016
  • Status: Terminated

Why It Was Recalled

Class II

Software defect:2D projection of contours in Verity. The defect causes structures to appear off from their true position in the DRR and the radiograph with an error that increases with the distance of the structure from isocenter

Record

Product
MEVION S250 Product Usage: Proton Radiation Therapy System
Quantity
6 units
Lots and codes
Serial Numbers: S250-0001 through S250-0006
Distribution
US Nationwide Distribution in the states of: FL, MO, NJ, OH, OK
Recalling firm
Mevion Medical Systems, Inc., Littleton, MA, United States
How the firm notified
Letter
FDA classified
Nov 14, 2016

Mevion Medical Systems, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.