Arca
Skip to content
Tools
Tools

MEVION S250i, MEVION S250; Proton Radiation Treatment System

Recall Z-2406-2020, part of FDA recall event 85748. See Mevion Medical Systems's company page.

Mevion Medical Systems, Inc. started this Class II device recall on May 15, 2020 (voluntary: firm initiated); FDA lists it as terminated, terminated Dec 11, 2020.

  • Classification: Class II
  • Started: May 15, 2020
  • Status: Terminated

Why It Was Recalled

Class II

Couch Correction moves sent after using a 3D CT scan are partially lost if the previous correction sent or recorded had some of the Couch Correction checkboxes turned off

Record

Product
MEVION S250i, MEVION S250; Proton Radiation Treatment System
Quantity
8 units
Lots and codes
Serial Numbers: MEVION S250-0001 MEVION S250-0002 MEVION S250-0003 MEVION S250-0004 MEVION S250-0005 MEVION S250-0006 MEVION S250-0007 MEVION S250-0008 UDI: (01)00864366000100 (11) 131209 (01)00864366000100 (11)150526 (01)00864366000100 (11)150113 (01)00864366000100 (11)150430 (01)00864366000100 (11)160127 (01)00864366000100 (11)160603 (01)00864366000124 (11) 180220 (01)00864366000124 (11) 181219
Distribution
Distribution to US in MO, NJ, OH, OK, FL, and DC, and the Netherlands
Recalling firm
Mevion Medical Systems, Inc., Littleton, MA, United States
How the firm notified
Letter
FDA classified
Jun 22, 2020

Mevion Medical Systems, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.