Alaris PCA Pump, Model 8120. The firm name on the label is CareFusion. The Alaris Syringe…

Recall Z-0323-2018, part of FDA recall event 78588. See CareFusion 303's company page.

CareFusion 303, Inc. started this Class II device recall on Nov 17, 2017 (voluntary: firm initiated); FDA lists it as terminated, terminated May 4, 2020.

  • Classification: Class II
  • Started: Nov 17, 2017
  • Status: Terminated
  • Products in this event: 2

Why It Was Recalled

Class II

The syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed.

Record

Product
Alaris PCA Pump, Model 8120. The firm name on the label is CareFusion. The Alaris Syringe Pump module and Alaris PCA module are infusion pump modules for the Alaris System. The syringe pump and PCA pump delivers medication or fluids in a manner similar to current syringe pumps on the market. The modules are indicated for use on adults, pediatrics and neonates for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous, subcutaneous or epidural. The Alaris Syringe pump and PCA modules can be connected to the Alaris PC unit which is the central programming, monitoring, and power supply component for the Alaris System. The syringe pump and PCA pump uses standard, plastic piston syringes and single-use administration sets with luer-lock connectors.
Quantity
73,611
Lots and codes
UDI 10885403812002. Product involved are those devices manufactured and/or serviced between 5/1/2013 through 4/30/2017.
Distribution
Worldwide Distribution - USA (nationwide) Distribution. There was also government and military distribution. Foreign distribution was made to Canada, United Arab Emirates, Bahrain, Canada, Switzerland, United Kingdom, Kuwait, Mexico, Malaysia, The Netherlands, New Zealand, Qatar, Saudi Arabia, Taiwan, and South Africa.
Recalling firm
CareFusion 303, Inc., San Diego, CA, United States
How the firm notified
Letter
FDA classified
Jan 8, 2018

Other Products in This Recall

Recall event 78588

CareFusion 303, Inc.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports