Arca
Skip to content
Tools
Tools

O-arm 1000 2nd Edition Imaging System

Recall Z-0278-2019, part of FDA recall event 81153. See Medtronic Navigation, Inc.-Littleton's company page.

Medtronic Navigation, Inc.-Littleton started this Class II device recall on Aug 30, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Jun 18, 2020.

  • Classification: Class II
  • Started: Aug 30, 2018
  • Status: Terminated
  • Products in this event: 10

Why It Was Recalled

Class II

Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.

Record

Product
O-arm 1000 2nd Edition Imaging System, BASE OARM Bl70000027230R SYSTEM 230V RWK, Model Number Bl70000027230R Product Usage: The O-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-arm Imaging System is compatible with certain Image Guided Surgery Systems.
Quantity
596 total
Lots and codes
All Serial Numbers
Distribution
Worldwide - US Distribution in the states: AK, AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and Puerto Rico. The products were distributed to the following foreign countries: Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kazakhstan, Korea, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zealand, Poland, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK, Venezuela.
Recalling firm
Medtronic Navigation, Inc.-Littleton, Littleton, MA, United States
How the firm notified
Letter
FDA classified
Oct 25, 2018

Other Products in This Recall