Arca
Skip to content
Tools
Tools

Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system:…

Recall Z-0404-2019, part of FDA recall event 81355. See Medtronic Navigation, Inc.-Littleton's company page.

Medtronic Navigation, Inc.-Littleton started this Class II device recall on Sep 17, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated May 28, 2020.

  • Classification: Class II
  • Started: Sep 17, 2018
  • Status: Terminated

Why It Was Recalled

Class II

O-arm 02 Imaging Systems correction to software version 4.1.0; new version of the user manual and an Image Acquisition System ("IAS") product label; replacing blown fuse of the new AC power input circuit

Record

Product
Medtronic Navigation 0-arm 02 Imaging System interventional fluoroscopic x-ray system: OARM BI70002000 SYS IMAGING 02 Model: BI70002000 OARM BI70002000R SYS IMAGING 02 REFURB Model: BI70002000R
Quantity
641 units
Lots and codes
Serial Numbers:C0875 through C2012
Distribution
Worldwide Distribution - US Nationwide Foreign: AUSTRALIA AUSTRIA BAHRAIN BANGLADESH BELGIUM BRAZIL CANADA CHILE COLOMBIA DENMARK FINLAND FRANCE GERMANY HR ICELAND INDIA IRELAND ISRAEL ITALY JAPAN JORDAN KOREA, REPUBLIC OF LUXEMBOURG MARTINIQUE MEXICO NETHERLANDS NEW ZEALAND POLAND PORTUGAL PUERTO RICO QATAR SAUDI ARABIA SINGAPORE SOUTH AFRICA SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED ARAB EMIRATES UNITED KINGDOM
Recalling firm
Medtronic Navigation, Inc.-Littleton, Littleton, MA, United States
How the firm notified
Letter
FDA classified
Nov 7, 2018

Medtronic Navigation, Inc.-Littleton's Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.