Omega Medical Imaging Elevating Monitor Suspension Systems
Recall Z-0160-2024, part of FDA recall event 93123. See Omega Medical Imaging's company page.
Omega Medical Imaging, Inc. started this Class II device recall on Sep 8, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Sep 8, 2023
- Status: Ongoing
An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable.
- Product
- Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085
- Quantity
- 31 units
- Lots and codes
- all units manufactured between 10-22-2012 and 07-09-2019
- Distribution
- Alabama, California, Illinois, Indiana, Kentucky, Louisiana, Missouri, Montana, North Carolina, Ohio, Oklahoma
- Recalling firm
- Omega Medical Imaging, Inc., Sanford, FL, United States
- How the firm notified
- FDA classified
- Oct 24, 2023
Newest first, one product per recall event; search them all
- Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085Z-0156-2024 · Class II · Feb 28, 2023 · Ongoing
- Omega Medical Imaging - 2 over 2 Elevating Monitor Suspension pillar/column accessory Device accessory that…Z-2621-2017 · Class II · Mar 23, 2017 · Terminated
- 2 over 2 Elevating Monitor Suspension - Omega Medical Imaging, LLC Elevating Monitor Suspension Systems.Z-2734-2016 · Class II · Jun 21, 2016 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports