Omega Medical Imaging Elevating Monitor Suspension
Recall Z-0156-2024, part of FDA recall event 93018. See Omega Medical Imaging's company page.
Omega Medical Imaging, Inc. started this Class II device recall on Feb 28, 2023 (voluntary: firm initiated); FDA lists it as ongoing.
- Classification: Class II
- Started: Feb 28, 2023
- Status: Ongoing
An actuator separated from the pivot mechanism on a lower monitor boom system column and became detached and was restrained by the installed safety cable at one location.
- Product
- Omega Medical Imaging Elevating Monitor Suspension, Part Number 1000-0085
- Quantity
- 52 units
- Lots and codes
- all units manufactured between 10-22-2012 and 07-09-2019
- Distribution
- Worldwide distribution - US Nationwide and the country of Pakistan.
- Recalling firm
- Omega Medical Imaging, Inc., Sanford, FL, United States
- How the firm notified
- FDA classified
- Oct 24, 2023
Newest first, one product per recall event; search them all
- Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085Z-0160-2024 · Class II · Sep 8, 2023 · Ongoing
- Omega Medical Imaging - 2 over 2 Elevating Monitor Suspension pillar/column accessory Device accessory that…Z-2621-2017 · Class II · Mar 23, 2017 · Terminated
- 2 over 2 Elevating Monitor Suspension - Omega Medical Imaging, LLC Elevating Monitor Suspension Systems.Z-2734-2016 · Class II · Jun 21, 2016 · Terminated