Arca
Skip to content
Tools
Tools

V-Twin, Catalog # 6002-800

Recall Z-0156-2020, part of FDA recall event 83582. See Vital Scientific's company page.

Vital Scientific N.V. started this Class II device recall on Jan 10, 2012 (voluntary: firm initiated); FDA lists it as terminated, terminated Apr 19, 2024.

  • Classification: Class II
  • Started: Jan 10, 2012
  • Status: Terminated
  • Products in this event: 5

Why It Was Recalled

Class II

The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.

Record

Product
V-Twin, Catalog # 6002-800, UDI # 03661540600180 Product Usage: automated chemistry analyzer intended for clinical use
Quantity
39 instruments
Lots and codes
serial numbers: 10-3331; 10-3332; 11-3301; 11-3304 through 11-3306; 11-3308 through 11-3312; 11-3315; 11-3316; 11-3318 through 11-3342; 12-3302.
Distribution
US Nationwide in the states of NY, and UT
Recalling firm
Vital Scientific N.V., Dieren, N/A, Netherlands
How the firm notified
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
FDA classified
Oct 16, 2019

Other Products in This Recall

Recall event 83582

Vital Scientific N.V.'s Other Recalls

Newest first, one product per recall event; search them all

FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.