V-Twin, Catalog # 6002-800
Recall Z-0038-2020, part of FDA recall event 83648. See Vital Scientific's company page.
Vital Scientific N.V. started this Class II device recall on Sep 7, 2018 (voluntary: firm initiated); FDA lists it as terminated, terminated Jun 29, 2022.
- Classification: Class II
- Started: Sep 7, 2018
- Status: Terminated
The action is being initiated because the cooling indicator board was improperly installed. This was first noticed by a service employee during assembly. If for some reason the cooling of the reagent rotor fails, the reagent cooling indicator light will not illuminate and the acoustic beeper will not go off. As a consequence the operator will not be warned and the reagent storage conditions may fall out of specification.
- Product
- V-Twin, Catalog # 6002-800, UDI: 03661540600180 - Product Usage: An in vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.
- Quantity
- 192 devices
- Lots and codes
- The following serial numbers: 14-3301-14-3344; 15-3301-15-3326; 16-3301 -16-3353; 17-3301 -17-3346; 18-3301-18-3323
- Distribution
- US Nationwide distribution including the state of New York.
- Recalling firm
- Vital Scientific N.V., Dieren, N/A, Netherlands
- How the firm notified
- FDA classified
- Oct 4, 2019
- V-Twin, REF 6002-800 or 6002-850, SN 18-3304, IVD, (01)03661540600180Z-0547-2020 · Class II · Nov 15, 2018 · Ongoing
- V-Twin, Model # 6002-800, UDI: 03661540600180 Product Usage: An in-vitro diagnostic device, which is an automated…Z-2558-2019 · Class II · Jan 18, 2016 · Ongoing
- Viva-E, Catalog # 6002-380, UDI # 03661540600050 Product Usage: automated chemistry analyzer intended for clinical…Z-0155-2020 · Class II · Jan 10, 2012 · Terminated
FDA's weekly Enforcement Report, as openFDA publishes it. A recall is classified by how likely the product is to cause harm: Class I is the most serious. openFDA device enforcement reports