Ipump Pain Management System, Model 2L3107K
Pump, Infusion. 510(k) K993387, from Baxter Healthcare Corp (Round Lake, IL). Product code FRN, class 2.
As of Nov 21, 2026, K993387 is an FDA 510(k) held by Baxter Healthcare Corp, substantially equivalent on Nov 1, 1999 after 24 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 24 days
- Recalls: 3
- Cited as a predicate by: 2 devices
- Bard Apii PumpK926385 · C.R. Bard, Inc. · Apr 18, 1994
- Patient Controlled Analgesia (Pca) Button CableK120209 · American I.V. Products, Inc. · Apr 12, 2012
- Ipump Pain Management System, Model 2L3107, 2L3107KK052973 · Baxter Healthcare Corp · Jan 25, 2007
| Dec 21, 2005 | Z-0551-06Delivery of unrequested PCA doses of medication due to damaged PCA cords, partial sticking of PCA button, and fluid ingress into the pump or PCA button. | Terminated |
| Oct 27, 2005 | Z-0549-06When infusing at flow rates below 0.5 mL/hr, the upstream occlusion alarm of the pumps may not detect the occlusion. | Terminated |
| Feb 10, 2005 | Z-0130-06Potential failures of the pump due to upstream occlusion alarms, electro-static discharge, system error 32 or 33, history printout errors and fluid volume in 'Use Previous Rx' when priming through the pump. | Terminated |
- RemunityPRO+ Pump for Remodulin® (treprostinil) InjectionK253703 · Deka R&D Corp · Jul 24, 2026
- FreedomEdge Infusion System; High-Flo SubQ Needle Set; Precision Flow Rate TubingK252015 · Koru Medical Systems, Inc. · Jan 28, 2026
- Spectrum IQ Infusion System with Dose IQ Safety Software (3570009)K251636 · Baxter Healthcare Corporation · Jul 28, 2025
- SIGMA Spectrum Infusion Pump with Master Drug LibraryK251640 · Baxter Healthcare Corporation · Jul 28, 2025
- MRidium 3870 MRI Infusion Pump System (3870)K242752 · Iradimed Corporation · May 28, 2025
- BD Alaris Infusion System with Guardrails Suite MXK243855 · Carefusion 303, Inc. · Apr 25, 2025
- Plum Solo™ Precision IV PumpK242114 · Icu Medical, Inc. · Apr 2, 2025
- Plum Duo™ Precision IV PumpK242115 · Icu Medical, Inc. · Apr 2, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Oct 8, 1999
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5725
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files