Nellcor Puritan Bennett, 760 Ventilator System

Ventilator, Continuous, Facility Use. 510(k) K984379, from Puritan Bennett Corp. (Galway, IE). Product code CBK, class 2.

As of Nov 20, 2026, K984379 is an FDA 510(k) held by Puritan Bennett Corp., substantially equivalent on Dec 10, 1998 after 3 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 3 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K984379, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jan 25, 2013Z-0821-2014, Z-0822-2014Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator…Terminated
Dec 17, 2004Z-0445-05Capacitor weakness in backup alarm system might fail causing lack of backup alarm to function in the event of total power loss.Terminated
Dec 16, 2004Z-0444-05Wires on battery may loosen causing disconnection.Terminated

Other CBK clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Special
Received
Dec 7, 1998
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files