Nellcor Puritan Bennett, 740 Ventilator System
Ventilator, Continuous, Facility Use. 510(k) K964540, from Puritan Bennett Corp. (Carlsbad, CA). Product code CBK, class 2.
As of Nov 18, 2026, K964540 is an FDA 510(k) held by Puritan Bennett Corp., substantially equivalent - subject to tracking reg. on Nov 14, 1997 after 367 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 367 days
- Recalls: 4
- Cited as a predicate by: 2 devices
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- 7200AE Microprocessor Ventilator W/Enhanced Hard.K902506 · Puritan Bennett Corp. · Jan 8, 1991
- Emerson 3MV Volume VentilatorK864394 · J. H. Emerson Co. · Dec 5, 1986
Cited as a predicate
2 later 510(k) summaries name this device as a predicate- Nellcor Puritan Bennett, 760 Ventilator SystemK984379 · Puritan Bennett Corp. · Dec 10, 1998
- Air Safety Hepa and Non-Hepa FiltersK033008 · Air Safety, Ltd. · May 13, 2004
Recalls
Recalls FDA ties to K964540, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jan 25, 2013 | Z-0821-2014, Z-0822-2014Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator… | Terminated |
| Dec 17, 2004 | Z-0445-05Capacitor weakness in backup alarm system might fail causing lack of backup alarm to function in the event of total power loss. | Terminated |
| Dec 16, 2004 | Z-0444-05Wires on battery may loosen causing disconnection. | Terminated |
Other CBK clearances
The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device- Evita (V800); Evita (V600)K253647 · Draegerwerk AG & CO Kgaa · May 7, 2026
- Inbentus VersatileK252663 · Iconic Solutions by Murcia, SL · Apr 14, 2026
- VPAP Pediatric Face MaskK251888 · Hsiner Co., Ltd. · Mar 12, 2026
- Vivo 45 LSK232365 · Oconnell Regulatory Consultants, Inc. · Dec 19, 2025
- F&P OptiNIV Hospital Vented Full Face Mask Compatible with Single-limb Circuits – Size A (ONIV117A); F&P OptiNIV…K252173 · Fisher & Paykel Healthcare, Ltd. · Sep 10, 2025
- SV600, SV800 VentilatorK243767 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 28, 2025
- AVEA disposable expiratory filter/water trap (11790)K251133 · Telesair, Inc. · May 8, 2025
- AF531 Oro-Nasal SE Face MaskK243394 · Respironics, Inc. · Dec 30, 2024
Record
- Type
- Traditional
- Received
- Nov 12, 1996
- Review panel
- Anesthesiology
- Regulation
- 21 CFR 868.5895
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files