Lifepak 500 Biphasic
Automated External Defibrillators (Non-Wearable). 510(k) K983393, from Physio-Control Corp. (Redmond, WA). Product code MKJ, class 3.
As of Nov 20, 2026, K983393 is an FDA 510(k) held by Physio-Control Corp., substantially equivalent - subject to tracking reg. on May 5, 1999 after 222 days, with 1 recall tied to it in FDA's recall database.
- Review time: 222 days
- Recalls: 1
- Cited as a predicate by: 5 devices
The LIFEPAK 500 AED is to be used only on patients in cardiopulmonary arrest. The patient must be unconscious, pulseless, and not breathing spontaneously before using the defibrillator to analyze the patient's ECG rhythm. It is not intended for use on children less than eight years of age, per ATT A guidelines.
- Lifepak 500K955854 · Physio-Control Corp. · Nov 4, 1996
- Personal DefibrillatorK832833 · Physio-Control, Inc. · Oct 26, 1984
- Fore Runner External Defibrillator/Defibrillator PadsK955628 · Heartstream, Inc. · Sep 10, 1996
- Heartstart 3000K896919 · Laerdal Medical Corp. · May 3, 1990
- Lifepak 12 BiphasicK991910 · Physio-Control Corp. · Sep 3, 1999
- Lifepak 500 Automated External DefibrillatorK052057 · Medtronic Emergency Response Systems, Inc. · Feb 17, 2006
- Lifepak, Model 1000K042404 · Medtronic Emergency Response Systems, Inc. · Dec 22, 2005
- Accessaed, Semi-Automatic External DefibrillatorsK020984 · Access Cardiosystems, Inc. · Aug 14, 2002
- Modification To: Lifepak 500 Automated External DefibrillatorK012428 · Medtronic Physio-Control Corp. · Sep 28, 2001
| Started | Recall | Status |
|---|---|---|
| Mar 20, 2020 | Z-1633-2020Automated External Defibrillators may not detect a patient connection due to mechanical wear-through of the contact plating on the therapy connector. The wear-through could result in the device not recognizing a patient… | Terminated |
- Type
- Traditional
- Received
- Sep 25, 1998
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.5310
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files