Arca
Skip to content
Tools
Tools

Accessaed, Semi-Automatic External Defibrillators

Automated External Defibrillators (Non-Wearable). 510(k) K020984, from Access Cardiosystems, Inc. (Concord, MA). Product code MKJ, class 3.

As of Nov 19, 2026, K020984 is an FDA 510(k) held by Access Cardiosystems, Inc., substantially equivalent - subject to tracking reg. on Aug 14, 2002 after 149 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 149 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The AccessAEDâ„¢ semi-automatic external defibrillator is intended to be used for the treatment of cardiac arrest. The AccessAED should be applied to victims who are unconscious, with absence of breathing, and absence of a detectable pulse. The AccessALSâ„¢ has the capability for manual operation by a user trained in advanced cardiac life support (ALS).

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K020984, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 3, 2004Z-0199-05Device may fail to deliver shock due to a faulty circuit boardTerminated
Dec 19, 2003Z-0356-04Difibrillator may fail to operate due to a faulty component on the circuit boardTerminated

Record

Type
Traditional
Received
Mar 18, 2002
Review panel
Cardiovascular
Regulation
21 CFR 870.5310

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files