Genesis Imrt Linear Accelerator
Accelerator, Linear, Medical. 510(k) K982502, from Siemens Medical Solutions USA, Inc. (Concord, CA). Product code IYE, class 2.
As of Nov 20, 2026, K982502 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Jul 31, 1998 after 11 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 11 days
- Recalls: 4
The GENESIS IMRT Linear Accelerator is intended to deliver megavoltage x-ray treatment for therapeutic application in the treatment of cancer. The GENESIS delivers these treatments conformally, shaping the field of radiation to the tumor volume. The depth of penetration emulating from the GENESIS ideally suits itself for treating deep-seated tumors while providing a skin sparing effect.
- Mevatron, M 2 ClassK882729 · Siemens Medical Laboratories, Inc. · Sep 19, 1988
| Dec 19, 2014 | Z-1059-2015A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal… | Terminated |
| May 28, 2014 | Z-2174-2014A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to… | Terminated |
| May 29, 2013 | Z-1663-2013The information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in… | Terminated |
| Nov 4, 2010 | Z-2639-2011The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis. | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jul 20, 1998
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files