Genesis Imrt Linear Accelerator

Accelerator, Linear, Medical. 510(k) K982502, from Siemens Medical Solutions USA, Inc. (Concord, CA). Product code IYE, class 2.

As of Nov 20, 2026, K982502 is an FDA 510(k) held by Siemens Medical Solutions USA, Inc., substantially equivalent on Jul 31, 1998 after 11 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 11 days
  • Recalls: 4

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The GENESIS IMRT Linear Accelerator is intended to deliver megavoltage x-ray treatment for therapeutic application in the treatment of cancer. The GENESIS delivers these treatments conformally, shaping the field of radiation to the tumor volume. The depth of penetration emulating from the GENESIS ideally suits itself for treating deep-seated tumors while providing a skin sparing effect.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K982502, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 19, 2014Z-1059-2015A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal…Terminated
May 28, 2014Z-2174-2014A safety risk exists with automatically sequenced treatment technique using the SIMTEC Auto Field Option where automatic movements of the gantry and the tabletop during an auto-sequenced treatment have the potential to…Terminated
May 29, 2013Z-1663-2013The information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in…Terminated
Nov 4, 2010Z-2639-2011The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Special
Received
Jul 20, 1998
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files