Mevatron, M 2 Class
Accelerator, Linear, Medical. 510(k) K882729, from Siemens Medical Laboratories, Inc. (Walnut Creek, CA). Product code IYE, class 2.
As of Nov 20, 2026, K882729 is an FDA 510(k) held by Siemens Medical Laboratories, Inc., substantially equivalent on Sep 19, 1988 after 75 days, with 25 recalls tied to it in FDA's recall database.
- Review time: 75 days
- Recalls: 25+
- Cited as a predicate by: 3 devices
- PrimusK993425 · Siemens Medical Solutions USA, Inc. · Nov 26, 1999
- MX MevatronK982513 · Siemens Medical Solutions USA, Inc. · Jul 31, 1998
- Genesis Imrt Linear AcceleratorK982502 · Siemens Medical Solutions USA, Inc. · Jul 31, 1998
| Feb 18, 2020 | Z-1678-2020, Z-1679-2020Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy. | Terminated |
| Jan 25, 2017 | Z-1436-2017Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to… | Terminated |
| Dec 1, 2011 | Z-0781-2012, Z-0798-2012, Z-0791-2012, Z-0787-2012, Z-0785-2012, Z-0788-2012, Z-0794-2012, Z-0786-2012, Z-0792-2012, Z-0796-2012, Z-0795-2012, Z-0783-2012, Z-0793-2012, Z-0789-2012, Z-0797-2012, Z-0780-2012, Z-0784-2012, Z-0790-2012Bolts holding the trim weight may break causing the plates of the counterweight to become loose and fall into the cover. | Terminated |
| Jan 11, 2008 | Z-1402-2008, Z-1401-2008Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact. | Terminated |
| Aug 7, 2007 | Z-0678-2008, Z-0682-2008Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in… | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Type
- Traditional
- Received
- Jul 6, 1988
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files