Synchron Lx 20 Clinical Chemistry System, Synchron Lx Albumin Reagent, Synchron Lx Creatinine Picric Reagent, Synchron L
Analyzer, Chemistry (Photometric, Discrete), For Clinical Use. 510(k) K965240, from Beckman Instruments, Inc. (Brea, CA). Product code JJE, class 1.
As of Nov 17, 2026, K965240 is an FDA 510(k) held by Beckman Instruments, Inc., substantially equivalent on Mar 10, 1997 after 70 days, with 12 recalls tied to it in FDA's recall database.
- Substantially equivalent: Mar 10, 1997
- Review time: 70 days
- Recalls: 12
- Cited as a predicate by: 4 devices
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Synchon Cx Delta Clin Syst (CX4, CX5, CX7 Delta)K950958 · Beckman Instruments, Inc. · May 8, 1995
- Synchron CX4/CX5 Clinical Chemistry Analyzer ReageK883181 · Beckman Instruments, Inc. · Sep 19, 1988
Cited as a Predicate
4 later 510(k) summaries name this device as a predicate- Synchron Cx Systems Lipase (Lipa) ReagentK970905 · Beckman Instruments, Inc. · Jun 4, 1997
- Synchron LX20 Clinical Systems, Creatinine Test SystemsK071277 · Beckman Coulter, Inc. · Jun 1, 2007
- Unicel Dxc 600 And 800 Synchron SystemsK042291 · Beckman Coulter, Inc. · Nov 12, 2004
- Modification To Synchron Lx Clinical Chemistry Systems (LX20 And LX20 Pro)K014034 · Beckman Coulter, Inc. · Dec 18, 2001
Recalls
Recalls FDA ties to K965240, newest first; each opens FDA's recall record| Apr 5, 2011 | Z-2130-2011The recall was initiated because Beckman Coulter has received reports of incorrect electrolyte results on Synchron LX systems due to a variety of maintenance-related hardware issues. | Terminated |
| Oct 22, 2010 | Z-2071-2011This recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified. | Terminated |
| Sep 22, 2010 | Z-1951-2013Beckman Coulter is recalling the Synchron Systems LX20/UniCel DxC ISE Reference Reagent because it has a potassium (K) response near the instrument calibration flag limits and may exceed the limits on some instruments. | Terminated |
| Aug 10, 2010 | Z-1619-2011The recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are… | Terminated |
| Aug 2, 2010 | Z-2389-2010The recall was initiated after Beckman Coulter confirmed reports that black / brown particulates were observed in some bottles of the Creatinine Alkaline Buffer (Buffer lot M911427 and M911518) in modular Creatinine… | Terminated |
| Apr 15, 2010 | Z-2388-2010UniCel DxC and Synchron LX instruments produced inaccurate (low) GLUCm results. False low results could affect or delay diagnosis or treatment. | Terminated |
| Apr 5, 2007 | Z-1094-2008Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall… | Terminated |
| Sep 6, 2005 | Z-0009-06Synchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour… | Terminated |
| Oct 10, 2003 | Z-1098-04Numeric results are erroneously accepted by the in-house information system without qualifying symbols "<" or ">". In addition, the in-house system may interpret the results incorrectly. | Terminated |
| Oct 7, 2003 | Z-1116-04'Initial Rate High' parameter for User Defined Reagents does not perform as indicated in the Synchron LX Clinical Systems Operations Manual. | Terminated |
| Sep 18, 2003 | Z-0096-04Sample wheel may home to the wrong position. | Terminated |
| Dec 16, 2002 | Z-0545-03The disk that controls the home position of the sample wheel comes loose preventing a proper home sequence. | Terminated |
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 30, 1996
- Review panel
- Clinical Chemistry
- Regulation
- 21 CFR 862.2160
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files