Arca
Skip to content

Intramedullary Nail System

Rod, Fixation, Intramedullary And Accessories. 510(k) K965098, from Zimmer, Inc. (Warsaw, IN). Product code HSB, class 2.

As of Nov 17, 2026, K965098 is an FDA 510(k) held by Zimmer, Inc., substantially equivalent on Feb 28, 1997 after 70 days, with 25 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 28, 1997
  • Review time: 70 days
  • Recalls: 25+
  • Cited as a predicate by: 7 devices

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K965098, newest first; the 25 newest; each opens FDA's recall record
Apr 28, 2017Z-2597-2017Zimmer Biomet is initiating a field action for sterile-packaged implants packaged in a single package configuration due to packaging design verification test failures.Terminated
Apr 28, 2017Z-2596-2017Zimmer Biomet is initiating a field action for sterile-packaged implants packaged in a single package configuration due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2558-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2572-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2560-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2551-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2570-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2550-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2554-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2571-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2561-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2567-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2553-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2559-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2557-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2569-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2549-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2552-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2568-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2556-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2573-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Feb 28, 2017Z-2555-2017Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.Terminated
Jan 11, 2016Z-1158-2016As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.Terminated
Jan 11, 2016Z-1140-2016As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.Terminated
Jan 11, 2016Z-1135-2016As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.Terminated

Other HSB Clearances

The newest 510(k)s for the same product code; their median review was 192 days; see every HSB device

Record

Type
Traditional
Received
Dec 20, 1996
Review panel
Orthopedic
Regulation
21 CFR 888.3020

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files